3M™ Attest™ Biological Indicator 1264 for Ethylene Oxide

  • Highly resistant, standardised population of Bacillus atrophaeus spores
  • Simple culturing, minimizes false positives due to contamination, cost effective
  • 48-hour standard readout
  • Complies with ISO 11138 Parts 1 and 2

Suggested Applications

  • Monitoring ethylene oxide (EO) sterilization processes

Description

The 3M™ Attest™ Biological Indicator 1264 (green cap) is designed for reliable monitoring for Ethylene Oxide sterilization process when used in conjunction with an appropriate Biological Indicator Incubator.

The 3M™ Attest™ 1264 Biological Indicator (green cap) is designed for monitoring of ethylene oxide (E0) sterilization processes. After sterilization, the vial is “crushed” to join the growth media with the processed spore strip. The BI is incubated for 48 hours for a visual color change readout. The presence of Bacillus atrophaeus spores is detected by a visual color change (media turns yellow). The yellow color change indicates a sterilization process failure. The final readout of a negative result (media remains green) is made after 48 hours of incubation.

Additional information

Biological Indicator System

Standard

Brand

Attest™

Model

1264

Product Color

Green

Product Type

Biological Indicator

Product Use

Load Monitoring

Sterilization Method

Ethylene Oxide

Units Per Pack

100

Reviews

There are no reviews yet.

Be the first to review “3M™ Attest™ Biological Indicator 1264 for Ethylene Oxide”

Your email address will not be published. Required fields are marked *